1. Visual Inspection
Each pump is examined for external damage, cracks, and signs of wear on the body and components. Cosmetic issues and potential safety concerns are noted for further action.
Fully Refurbished
1-Year Warranty
Biomed Tested & Calibrated
Complete Working System
Refurbished Alaris MedSystem III 2863 Infusion IV Pump
Refurbished • 1-Year Warranty • Compact three-channel MedSystem III volumetric infusion pump
The Alaris MedSystem III 2863 is a compact multi-channel volumetric infusion pump with three independent fluid-delivery channels in one device. Each channel can be programmed separately, allowing multiple infusions to be managed from a single pump while reducing equipment footprint at the bedside or during patient transport.
Fully refurbished, biomed tested and calibrated by United Infusion, and backed by a 1-year warranty, the MedSystem III 2863 provides three independently programmable infusion channels in a portable unit weighing approximately 5.1 pounds with the pole clamp.
The pump supports primary and secondary infusion programming as well as a Dose Rate Calculator that determines volumetric or dose rates from patient weight, drug concentration, diluent volume, and dosing parameters. Six configurable Device Types allow parameters such as rate, volume, pressure, KVO, and air-in-line thresholds to be adapted for different clinical applications.
Model 2863 is specifically distinguished by its standard 4-pin locking AC-adapter connection. Model 2860 uses a detachable power connector instead.
| Manufacturer | Alaris Medical Systems / IVAC / CareFusion |
| Model | 2863 — MedSystem III |
| Type | Three-channel volumetric multi-channel infusion pump |
| Infusion Channels | 3 independent channels |
| Delivery Routes | Intravenous, intra-arterial, subcutaneous, and epidural when used in accordance with manufacturer instructions |
| Dimensions | 7.875″ H × 6″ W × 2.10″ D (20.00 × 15.24 × 5.33 cm) |
| Weight | Approximately 5.1 lb (2.32 kg), including pole clamp |
| Flow Rate Range | 0.1–999 mL/h per channel |
| Volume Range | 0.1–9999 mL per channel |
| KVO Rate Range | 0.1–20.0 mL/h; configured KVO may vary according to Device Type and institutional configuration |
| System Accuracy | ±5% from 1.0–999 mL/h and ±10% from 0.1–0.9 mL/h under manufacturer-specified test conditions |
| Primary Infusion | Each channel can be programmed for a specified delivery rate or delivery period |
| Secondary Infusion | Supports sequential delivery at a separately programmed secondary rate; when secondary volume remaining reaches zero, primary infusion resumes automatically |
| Dose Rate Calculator | Calculates volumetric or dose rate using patient weight, drug concentration, diluent volume, and dosing parameters |
| Drug Programming | Supports selection of listed generic drug names and the DRUG? option for a drug not listed in the pump or for an alternative dosing regimen |
| Device Types | General Purpose, Neonatal, Controller Pressure, Operating Room, General Purpose II, and Operating Room II |
| Air-in-Line Detection | Air-in-line sensor on each channel; factory/default threshold is 500 µL except Neonatal Device Type at 50 µL, subject to configuration |
| Occlusion Pressure | Default maximum 15 psi for most Device Types; Controller Pressure uses a 3 ft H₂O setting. Parameters may be reconfigured by qualified personnel. |
| Administration Sets | Use only approved IVAC MedSystem III administration sets; manufacturer Directions for Use specifies IVAC 28 and 25 Series sets |
| Free-Flow Protection | IVAC 28 and 25 Series set cassettes use a slide clamp that occludes the tubing when fully extended during cassette removal |
| Communications | COMM receptacle supports RS-232-compatible communications for appropriately configured monitoring, configuration, and maintenance systems |
| Main Battery | Rechargeable NiCd battery pack |
| Memory Backup Battery | Nonrechargeable lithium battery |
| Battery Runtime | Minimum approximately 6 hours with all three channels operating at 125 mL/h and backlight usage of approximately 2 minutes per hour, using a fully charged battery under manufacturer-specified conditions |
| Power Consumption | 6 W AC power through an approved MedSystem III AC adapter |
| Compatible AC Adapters | IVAC MedSystem III Model 1555 or Model 1550 |
| Model 1555 Output | 7.5 VDC at 1 A with approximately 10 ft (3.05 m) cord |
| Model 1550 Output | 8.5 VDC at 750 mA with approximately 8.5 ft (2.59 m) cord |
| Model 2863 Power Connection | 4-pin locking AC-adapter connector |
| Operating Temperature | 50–104°F (10–40°C) |
| Storage Temperature | Below 95°F (35°C) recommended for optimum battery life |
| Maximum Storage Temperature | 131°F (55°C) |
| Case Material | Impact-resistant polycarbonate / ABS alloy |
| Configuration | Device Type parameters, rate and volume limits, pressure settings, KVO, air-in-line thresholds, communications functions, and other options may vary by individual pump configuration |
| Condition | Refurbished |
| Warranty | 1-Year United Infusion warranty |
Download the Alaris MedSystem III Directions for Use for operating instructions, primary and secondary programming, Dose Rate Calculator operation, Device Types, alarms, configuration, technical specifications, maintenance, and safety precautions.
Fully Refurbished
1-Year Warranty
Biomed Tested & Calibrated
Complete Working System
Refurbished Alaris MedSystem III 2863 Infusion IV Pump
Refurbished • 1-Year Warranty • Compact three-channel MedSystem III volumetric infusion pump
The Alaris MedSystem III 2863 is a compact multi-channel volumetric infusion pump with three independent fluid-delivery channels in one device. Each channel can be programmed separately, allowing multiple infusions to be managed from a single pump while reducing equipment footprint at the bedside or during patient transport.
Fully refurbished, biomed tested and calibrated by United Infusion, and backed by a 1-year warranty, the MedSystem III 2863 provides three independently programmable infusion channels in a portable unit weighing approximately 5.1 pounds with the pole clamp.
The pump supports primary and secondary infusion programming as well as a Dose Rate Calculator that determines volumetric or dose rates from patient weight, drug concentration, diluent volume, and dosing parameters. Six configurable Device Types allow parameters such as rate, volume, pressure, KVO, and air-in-line thresholds to be adapted for different clinical applications.
Model 2863 is specifically distinguished by its standard 4-pin locking AC-adapter connection. Model 2860 uses a detachable power connector instead.
| Manufacturer | Alaris Medical Systems / IVAC / CareFusion |
| Model | 2863 — MedSystem III |
| Type | Three-channel volumetric multi-channel infusion pump |
| Infusion Channels | 3 independent channels |
| Delivery Routes | Intravenous, intra-arterial, subcutaneous, and epidural when used in accordance with manufacturer instructions |
| Dimensions | 7.875″ H × 6″ W × 2.10″ D (20.00 × 15.24 × 5.33 cm) |
| Weight | Approximately 5.1 lb (2.32 kg), including pole clamp |
| Flow Rate Range | 0.1–999 mL/h per channel |
| Volume Range | 0.1–9999 mL per channel |
| KVO Rate Range | 0.1–20.0 mL/h; configured KVO may vary according to Device Type and institutional configuration |
| System Accuracy | ±5% from 1.0–999 mL/h and ±10% from 0.1–0.9 mL/h under manufacturer-specified test conditions |
| Primary Infusion | Each channel can be programmed for a specified delivery rate or delivery period |
| Secondary Infusion | Supports sequential delivery at a separately programmed secondary rate; when secondary volume remaining reaches zero, primary infusion resumes automatically |
| Dose Rate Calculator | Calculates volumetric or dose rate using patient weight, drug concentration, diluent volume, and dosing parameters |
| Drug Programming | Supports selection of listed generic drug names and the DRUG? option for a drug not listed in the pump or for an alternative dosing regimen |
| Device Types | General Purpose, Neonatal, Controller Pressure, Operating Room, General Purpose II, and Operating Room II |
| Air-in-Line Detection | Air-in-line sensor on each channel; factory/default threshold is 500 µL except Neonatal Device Type at 50 µL, subject to configuration |
| Occlusion Pressure | Default maximum 15 psi for most Device Types; Controller Pressure uses a 3 ft H₂O setting. Parameters may be reconfigured by qualified personnel. |
| Administration Sets | Use only approved IVAC MedSystem III administration sets; manufacturer Directions for Use specifies IVAC 28 and 25 Series sets |
| Free-Flow Protection | IVAC 28 and 25 Series set cassettes use a slide clamp that occludes the tubing when fully extended during cassette removal |
| Communications | COMM receptacle supports RS-232-compatible communications for appropriately configured monitoring, configuration, and maintenance systems |
| Main Battery | Rechargeable NiCd battery pack |
| Memory Backup Battery | Nonrechargeable lithium battery |
| Battery Runtime | Minimum approximately 6 hours with all three channels operating at 125 mL/h and backlight usage of approximately 2 minutes per hour, using a fully charged battery under manufacturer-specified conditions |
| Power Consumption | 6 W AC power through an approved MedSystem III AC adapter |
| Compatible AC Adapters | IVAC MedSystem III Model 1555 or Model 1550 |
| Model 1555 Output | 7.5 VDC at 1 A with approximately 10 ft (3.05 m) cord |
| Model 1550 Output | 8.5 VDC at 750 mA with approximately 8.5 ft (2.59 m) cord |
| Model 2863 Power Connection | 4-pin locking AC-adapter connector |
| Operating Temperature | 50–104°F (10–40°C) |
| Storage Temperature | Below 95°F (35°C) recommended for optimum battery life |
| Maximum Storage Temperature | 131°F (55°C) |
| Case Material | Impact-resistant polycarbonate / ABS alloy |
| Configuration | Device Type parameters, rate and volume limits, pressure settings, KVO, air-in-line thresholds, communications functions, and other options may vary by individual pump configuration |
| Condition | Refurbished |
| Warranty | 1-Year United Infusion warranty |
Download the Alaris MedSystem III Directions for Use for operating instructions, primary and secondary programming, Dose Rate Calculator operation, Device Types, alarms, configuration, technical specifications, maintenance, and safety precautions.
Refurbished Alaris MedSystem III 2863 Infusion IV Pump
Refurbished • 1-Year Warranty • Compact three-channel MedSystem III volumetric infusion pump
The Alaris MedSystem III 2863 is a compact multi-channel volumetric infusion pump with three independent fluid-delivery channels in one device. Each channel can be programmed separately, allowing multiple infusions to be managed from a single pump while reducing equipment footprint at the bedside or during patient transport.
Fully refurbished, biomed tested and calibrated by United Infusion, and backed by a 1-year warranty, the MedSystem III 2863 provides three independently programmable infusion channels in a portable unit weighing approximately 5.1 pounds with the pole clamp.
The pump supports primary and secondary infusion programming as well as a Dose Rate Calculator that determines volumetric or dose rates from patient weight, drug concentration, diluent volume, and dosing parameters. Six configurable Device Types allow parameters such as rate, volume, pressure, KVO, and air-in-line thresholds to be adapted for different clinical applications.
Model 2863 is specifically distinguished by its standard 4-pin locking AC-adapter connection. Model 2860 uses a detachable power connector instead.
| Manufacturer | Alaris Medical Systems / IVAC / CareFusion |
| Model | 2863 — MedSystem III |
| Type | Three-channel volumetric multi-channel infusion pump |
| Infusion Channels | 3 independent channels |
| Delivery Routes | Intravenous, intra-arterial, subcutaneous, and epidural when used in accordance with manufacturer instructions |
| Dimensions | 7.875″ H × 6″ W × 2.10″ D (20.00 × 15.24 × 5.33 cm) |
| Weight | Approximately 5.1 lb (2.32 kg), including pole clamp |
| Flow Rate Range | 0.1–999 mL/h per channel |
| Volume Range | 0.1–9999 mL per channel |
| KVO Rate Range | 0.1–20.0 mL/h; configured KVO may vary according to Device Type and institutional configuration |
| System Accuracy | ±5% from 1.0–999 mL/h and ±10% from 0.1–0.9 mL/h under manufacturer-specified test conditions |
| Primary Infusion | Each channel can be programmed for a specified delivery rate or delivery period |
| Secondary Infusion | Supports sequential delivery at a separately programmed secondary rate; when secondary volume remaining reaches zero, primary infusion resumes automatically |
| Dose Rate Calculator | Calculates volumetric or dose rate using patient weight, drug concentration, diluent volume, and dosing parameters |
| Drug Programming | Supports selection of listed generic drug names and the DRUG? option for a drug not listed in the pump or for an alternative dosing regimen |
| Device Types | General Purpose, Neonatal, Controller Pressure, Operating Room, General Purpose II, and Operating Room II |
| Air-in-Line Detection | Air-in-line sensor on each channel; factory/default threshold is 500 µL except Neonatal Device Type at 50 µL, subject to configuration |
| Occlusion Pressure | Default maximum 15 psi for most Device Types; Controller Pressure uses a 3 ft H₂O setting. Parameters may be reconfigured by qualified personnel. |
| Administration Sets | Use only approved IVAC MedSystem III administration sets; manufacturer Directions for Use specifies IVAC 28 and 25 Series sets |
| Free-Flow Protection | IVAC 28 and 25 Series set cassettes use a slide clamp that occludes the tubing when fully extended during cassette removal |
| Communications | COMM receptacle supports RS-232-compatible communications for appropriately configured monitoring, configuration, and maintenance systems |
| Main Battery | Rechargeable NiCd battery pack |
| Memory Backup Battery | Nonrechargeable lithium battery |
| Battery Runtime | Minimum approximately 6 hours with all three channels operating at 125 mL/h and backlight usage of approximately 2 minutes per hour, using a fully charged battery under manufacturer-specified conditions |
| Power Consumption | 6 W AC power through an approved MedSystem III AC adapter |
| Compatible AC Adapters | IVAC MedSystem III Model 1555 or Model 1550 |
| Model 1555 Output | 7.5 VDC at 1 A with approximately 10 ft (3.05 m) cord |
| Model 1550 Output | 8.5 VDC at 750 mA with approximately 8.5 ft (2.59 m) cord |
| Model 2863 Power Connection | 4-pin locking AC-adapter connector |
| Operating Temperature | 50–104°F (10–40°C) |
| Storage Temperature | Below 95°F (35°C) recommended for optimum battery life |
| Maximum Storage Temperature | 131°F (55°C) |
| Case Material | Impact-resistant polycarbonate / ABS alloy |
| Configuration | Device Type parameters, rate and volume limits, pressure settings, KVO, air-in-line thresholds, communications functions, and other options may vary by individual pump configuration |
| Condition | Refurbished |
| Warranty | 1-Year United Infusion warranty |
Download the Alaris MedSystem III Directions for Use for operating instructions, primary and secondary programming, Dose Rate Calculator operation, Device Types, alarms, configuration, technical specifications, maintenance, and safety precautions.
At United Infusion, we follow a comprehensive refurbishing process for infusion pumps to help ensure they operate safely and reliably. This process follows manufacturer guidelines and sanitation standards to meet high quality expectations.
Below are the key steps of our refurbishing process:
Each pump is examined for external damage, cracks, and signs of wear on the body and components. Cosmetic issues and potential safety concerns are noted for further action.
The pump is carefully disassembled to access internal components, including the motor, gears, tubing pathways, and electronics. All parts are handled with care to avoid added damage.
Appropriate disinfectants are used to thoroughly clean all parts and remove residual fluids, blood, or debris. This step follows sanitation standards to help ensure the device is free from contaminants.
Individual parts, including the motor, battery, display screen, buttons, and connectors, are inspected for wear, damage, or malfunction. Worn or damaged components are replaced with manufacturer-approved parts.
The pump is adjusted to support accurate fluid delivery rates using standardized testing methods, and calibration is verified against manufacturer specifications. The latest compatible software version is then installed to improve functionality.
The pump goes through functional testing, including alarms, flow rate accuracy, occlusion detection, and bolus delivery. It is then reassembled, connections are checked, and a final visual and functional inspection is completed to confirm it meets safety standards and customer requirements.
"We needed our syringe pumps serviced for scheduled maintenance to continue to provide safe care. In the past, we had to send them off with a 6 weeks turn around time. Then we found United Infusion. 30 min down the road, 24 hr turn around and 30 minutes down the road. We are extremely grateful!!!"
United Infusion supports equipment sales, repair, and rentals. Our team can help match the right unit, answer product questions, and support your equipment needs.